The approach for treatment of large and fusiform intracranial aneurysms has evolved from stent-assisted coiling to treatment with flow-diverting stents. The treatment results for these stents are promising; however, early postprocedural aneurysm rupture has been described. The exact cause of rupture is unknown but might be related to intra-aneurysmal flow and pressure changes. We measured intra-aneurysmal pressure before, during, and after placement of a flow-diverting stent by using a dual-sensor guidewire. The pressure inside the aneurysm momentarily decreased during placement but was restored to baseline values within minutes. The flow-diverting stent does not seem to protect the aneurysm from the stress induced by pressure or pressure changes within the lumen.
机译:用于治疗大型和梭形颅内动脉瘤的方法已经从支架辅助卷绕到流动转移支架处理。这些支架的治疗结果很有前景;然而,已经描述了早期的后期动脉瘤破裂。断裂的确切原因是未知的,但可能与内胸腔内流动和压力变化有关。我们通过使用双传感器导丝在放置流动转移支架之前,期间和之后测量动脉瘤内压力。在放置期间动脉瘤内部的压力瞬间下降,但在几分钟内恢复到基线值。流动转移支架似乎似乎不会保护动脉瘤免受腔内压力或压力变化引起的应力。