首页> 美国卫生研究院文献>other >FLOW DYNAMIC COMPARISON BETWEEN RECESSED HINGE AND OPEN PIVOT BI-LEAFLET HEART VALVE DESIGNS
【2h】

FLOW DYNAMIC COMPARISON BETWEEN RECESSED HINGE AND OPEN PIVOT BI-LEAFLET HEART VALVE DESIGNS

机译:嵌入式铰链与开放式枢轴双传单心阀设计的流动性比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The flow dynamics through the peripheral and hinge regions of a bi-leaflet mechanical heart valve are complex and result in abnormally high shear stresses particularly during the closing phase of the valve function. It has been observed that, the late stages of closure is more significant in the dynamics of platelet activation; therefore, the later stages of closure is simulated by solving the two-dimensional Navier–Stokes equations using an Eulerian Levelset based sharp interface Cartesian grid method. Using a fixed Cartesian mesh incorporating local mesh refinement for solution accuracy and efficiency, the flow through and within a recessed hinge design and an open pivot hinge design is compared. Platelets are modelled as point particles by Lagrangian particle tracking algorithm with one way coupling. A dilute particle flow is assumed and particle-particle interactions are neglected. It was observed that the hinge region of the open pivot valve indicated a lower potential for activation of platelets compared to that in valves with a recessed hinge design.