An interlaboratory comparison of bone lead measurements via K-shell X-ray fluorescence spectrometry: validation against inductively coupled plasma mass spectrometry

机译:经由K-壳X射线荧光光谱法骨铅测量的实验室间比较:对电感耦合等离子体质谱法确认

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

109Cd-based K-shell X-ray fluorescence spectrometry (hereafter, for brevity, XRF) is used, often in epidemiological studies, to perform non-invasive, in vivo measurements of lead in bone. We conducted the first interlaboratory study of XRF via the circulation of nine goat tibiæ in which the mean lead value ranged from 4.0 µg g−1 to 55.3 µg g−1 bone mineral. The test tibiæ were subsequently analyzed via nitric acid digestion followed by lead determination by inductively coupled plasma mass spectrometry (ICP-MS) – along with certified reference materials for bone lead – thus providing measurement traceability to SI units. Analysis of dried bone for lead via nitric acid digestion and ICP-MS yields mass fraction data in units of µg g−1 dry weight. The mean bone lead value based on ICP-MS analysis ranged from 1.8 µg g−1 to 35.8 µg g−1 dry weight. For comparison purposes, XRF-measured Pb values (µg g−1 bone mineral) were converted into the ICP-MS-measured units (µg g−1dry weight bone) by multiplying the former by the average ash fraction from the nine tibiæ. Eight of the XRF systems did not yield a significant bias for any of the nine tibiæ; one system was biased for one of the tibiæ; two systems were biased for two tibiæ; one system was biased for four tibiæ; two systems (813-1 and 804-2) were biased for five tibiæ and one system (801-1) was biased for six of the nine tibiæ. Average bias for the systems (under those particular operating conditions) that were biased for the majority of samples ranged from −2.6 µg g−1 (−15.7%) to 5.1 µg g−1 (30.7%) dry weight bone. All participants now have the ICP-MS data, allowing any corrective actions deemed necessary to be implemented. The ICP-MS data, however, indicated that the lead mass fraction varied considerably with the sampling location within the tibiæ, to the extent of exceeding XRF variability for the higher lead values. Material heterogeneity is an unavoidable reality of measuring lead in bone.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号