An Intrathecally Located Broken Catheter Used for an Intrathecal Drug Delivery System

机译:用于鞘内给药系统的位于鞘内的折断导管

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The intrathecal drug delivery system (ITDDS), an effective treatment tool for intractable spasticity and pain, is associated with various complications but breakage of the catheter is rare. We report the case of a 50-yr-old man with ITDDS, in whom an intrathecal catheter was severed, resulting in a 28.6-cm-long intrathecal fragment. The catheter completely retracted into the intrathecal space from the anchor site. The catheter was severed during spine flexion, and the total distal fragment was repositioned in the intrathecal space. Although the outcome of ITDDS was associated with the length or diameter of the broken catheter, no neurologic complications occurred in our patient. Thus, we inserted another catheter instead of removing the old one. Thereafter, the patient has been regularly followed up, and no neurologic complications have developed during the 28 months.
机译:鞘内给药系统(ITDDS)是一种治疗顽固性痉挛和疼痛的有效治疗工具,具有多种并发症,但导管断裂很少见。我们报道了一个50岁的ITDDS患者的案例,其中鞘内导管被切断,导致28.6厘米长的鞘内碎片。导管从锚定位置完全缩回鞘内空间。在脊柱屈曲过程中,导管被切断,总的远端碎片被重新放置在鞘内空间中。尽管ITDDS的结果与导管断裂的长度或直径有关,但在我们的患者中未发生神经系统并发症。因此,我们插入了另一根导管,而不是卸下旧导管。此后,对该患者进行了定期随访,在28个月内未发生神经系统并发症。