首页> 美国卫生研究院文献>Sensors (Basel Switzerland) >An EEG Experimental Study Evaluating the Performance of Texas Instruments ADS1299
【2h】

An EEG Experimental Study Evaluating the Performance of Texas Instruments ADS1299

机译:脑电图实验研究评估德州仪器(TI)ADS1299的性能

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Texas Instruments ADS1299 is an attractive choice for low cost electroencephalography (EEG) devices owing to its low power consumption and low input referred noise. To date, there have been no rigorous evaluations of its performance. In this EEG experimental study we evaluated the performance of the ADS1299 against a high quality laboratory-based system. Two self-paced lower limb motor tasks were performed by 22 healthy participants. Recorded power across delta, theta, alpha, and beta EEG bands, the power ratio across the motor tasks, pre-movement noise, and signal-to-noise ratio were obtained for evaluation. The amplitude and time of the negative peak in the movement-related cortical potentials (MRCPs) extracted from the EEG data were also obtained. Using linear mixed models, no statistically significant differences (p > 0.05) were found in any of these measures across the two systems. These findings were further supported by evaluation of cosine similarity, waveform differences, and topographic maps. There were statistically significant differences in MRCPs across the motor tasks in both systems. We conclude that the performance of the ADS1299 in combination with wet Ag/AgCl electrodes is analogous to that of a laboratory-based system in a low frequency (<40 Hz) EEG recording.
机译:德州仪器(TI)ADS1299由于其低功耗和低输入参考噪声而成为低成本脑电图(EEG)设备的诱人选择。迄今为止,尚未对其性能进行严格的评估。在这项EEG实验研究中,我们针对基于高质量实验室的系统评估了ADS1299的性能。 22名健康参与者完成了两项自定进度的下肢运动任务。记录了跨δ,θ,α和βEEG频段记录的功率,运动任务的功率比,运动前的噪声和信噪比,以进行评估。还获得了从EEG数据中提取的运动相关皮层电位(MRCP)中负峰的幅度和时间。使用线性混合模型,在两个系统中的任何这些度量中均未发现统计学上的显着差异(p> 0.05)。余弦相似性,波形差异和地形图的评估进一步支持了这些发现。在两个系统中,跨整个运动任务的MRCP都有统计学上的显着差异。我们得出的结论是,ADS1299与湿式Ag / AgCl电极结合使用的性能类似于基于实验室的系统在低频(<40 Hz)脑电图记录中的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号