Repeatability Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices

机译:两种非接触式测厚仪测量中央角膜厚度的可重复性可重复性和一致性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This study was designed to assess the repeatability, reproducibility, and agreement of Noncontact Pachymetry (NPC) (Topcan TRK-2P) and the non-contact optical low coherence reflectometer (Lenstar LS 900) devices for measuring Central Corneal Thickness (CCT) of healthy corneas. A total of 82 healthy volunteers were evaluated. The first observer used both the TopconTRK-2P and Lenstar 900 devices while the second observer only used the TopconTRK-2P for the measurements. The measurements with either device were repeated three times for each patient, consecutively. The central corneal thickness measurements with the Topcon TRK-2P revealed mean ± Standard Deviation (SD) values of 553.1 ± 36.1 micrometer (µm) for the first observer and 552.3 ± 35.9µm for the second observer and the mean ± SD of CCT was 537.3 ± 34.8µm with the Lenstar 900. The difference between the CCT measurements of the observers using the Topcon TRK-2P (P = 0.142) was insignificant. However, significantly lower measurements were found with the Lenstar 900 compared with the Topcon TRK-2P (P ˂ 0.001). The central corneal thickness measurements obtained by the Topcon TRK-2P were found to have high repeatability for both observers with a lower SD, less than 1% Coefficient of Variation (CV) and higher than 0.99 Intra-Class Correlation Coefficient (ICC) (Observer 1: 3.77 SD, 0.68 CV and 0.995 ICC; the second observer: 3.58 SD, 0.65 CV and 0.995 ICC). There was an excellent inter-observer reproducibility between the two observers for Topcon TRK-2P with 2.71 SD, 0.49 CV, and 0.994 ICC. The Bland-Altman plot showed high agreement between the two devices. These results suggest that the TopconTRK-2P is a reliable device for evaluating CCT in healthy corneas compared with Lenstar 900.
机译:本研究旨在评估非接触式测厚仪(NPC)(Topcan TRK-2P)和非接触式光学低相干反射仪(Lenstar LS 900)的可重复性,可重复性和一致性,该设备用于测量健康人的中央角膜厚度(CCT)角膜。总共对82名健康志愿者进行了评估。第一个观察者同时使用TopconTRK-2P和Lenstar 900设备,而第二个观察者仅使用TopconTRK-2P进行测量。对于每位患者,使用任一设备进行的测量均连续重复三次。使用Topcon TRK-2P进行的中央角膜厚度测量显示,第一位观察者的平均±标准偏差(SD)值为553.1±36.1微米(µm),第二位观察者的平均±SD偏差为552.3±35.9µm,CCT的平均±SD为537.3 Lenstar 900的测量精度为±34.8µm。使用Topcon TRK-2P的观察者的CCT测量值之间的差异不明显(P = 0.142)。但是,与Topcon TRK-2P相比,Lenstar 900的测量值明显更低(P≤0.001)。发现拓普康TRK-2P获得的中央角膜厚度测量值对于具有较低SD,小于1%变异系数(CV)和大于0.99类内相关系数(ICC)的观察者均具有很高的重复性(观察者) 1:3.77 SD,0.68 CV​​和0.995 ICC;第二个观察者:3.58 SD,0.65 CV和0.995 ICC)。 Topcon TRK-2P的两个观察者之间具有出色的观察者间可重复性,具有2.71 SD,0.49 CV和0.994 ICC。 Bland-Altman图显示了两种设备之间的高度一致性。这些结果表明,与Lenstar 900相比,TopconTRK-2P是评估健康角膜CCT的可靠设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号