Determining the accuracy of an eye-tracking system for laser refractive surgery

机译:确定用于激光屈光手术的眼动追踪系统的准确性

获取原文
获取原文并翻译 | 示例

摘要

The accuracy of an eye tracking system designed for laser refractive surgery is determined. An enucleated porcine eye is attached to a scanning device and the movement measured using the eye tracking system. The recorded position is compared to the preprogrammed position of the scanning device. The system demonstrated an accuracy of 0.06mm for an intact cornea and 0.1mm for a cornea with a thin flap removed. The average decentration in laser refractive surgery is 0.4mm, so the implementation of this eye tracker would lead to improved alignment between the laser and eye during laser refractive surgery.
机译:确定了为激光屈光手术设计的眼动追踪系统的准确性。将去核猪眼固定在扫描设备上,并使用眼动跟踪系统测量其运动。将记录的位置与扫描设备的预编程位置进行比较。该系统显示出完整角膜的精度为0.06mm,角膜去除了薄瓣的精度为0.1mm。激光屈光手术的平均偏心为0.4毫米,因此实施该眼动仪将改善激光屈光手术期间激光与眼睛之间的对准度。