Gas Ultrasonic meter (GUSM) can be calibrated with good repeatability and accuracy using Master meter series validation. On the design aspect, all meter-runs should be designed to have a minimum 3 runs (2 operations + 1 spare run which is also the master meter-run) instead of 2 runs (one operation + 1 as spare which is also the MM). With 2 meter-runs system, master meter series validation can only be done at operating flowrate conditions. However with 3 meter-runs system, adjustment of flowrate (full range) is possible by manipulating valves in the system, as has been achieved in Shell Sarawak. Before removal of master USM for full range 6 points calibration at 3rd party calibration facility, a full range 6 points master meter series validation should be carried out for the operating meter runs (USM).
机译:使用主仪表系列验证,可以以良好的重复性和准确性校准气体超声计(GUSM)。在设计方面,所有仪表运行都应该设计为具有最小3个运行(2操作+ 1备用运行,也是主计的运行)而不是2个运行(一个操作+ 1作为MM也是MM的备用)。使用2米运行的系统,主仪表系列验证只能在运行流量条件下进行。然而,对于3米运行的系统,通过在系统中操纵阀门可以进行流量(全系列)的调整,如壳牌沙捞越中的实现。在拆除母版USM进行全范围6点校准3分校准设施中,应为运行仪表(USM)进行全系列6点主仪表系列验证。