COMPARISON OF pH VALUES OF PRIMARY BUFFER MATERIALS

机译:原发性缓冲材料pH值的比较

获取原文

摘要

The required comparability of pH values can only be ensured if the reference materials used for the calibration of pH measurement devices are traceable to a common pH scale. The multi-primary standard scale most frequently used at present and recommended by OIML [3] and also as one choice by IUPAC [2] is provided by a number of fixed points. These fixed points are pH values of buffer solutions of internationally agreed nominal composition, measured using an electrochemical cell without liquid junction. Uncertainties due to diffusion potentials are therefore excluded. With the aim of improving the uniformity of pH measurements in Europe, a first European comparison of primary pH measurement procedures, the EUROMET project 300, took place in 1993/94. The project showed good agreement between the pH results of the participating laboratories, except in one case, namely that of the phthalate buffer [1]. To continue this work, a second EUROMET project 370 "Intercomparison of standard measurement devices for pH" was carried out in 1996 [4]. This project, in which comparisons on four buffer materials have been carried out, is described in this paper. The pH of the phthalate buffer material was again measured. Apart from the partners of the EUROMET project 370 laboratories from Portugal and Great Britain joined this comparison. The partners from Portugal and Great Britain did not have a documented quality assurance system, so they do not fulfil the requirements for EUROMET projects. These participants use the same measurement procedure and the same materials and they have experience in high-precision pH measurements. For these reasons, their results are included in the report presented here. All partners are listed in Table 1.
机译:只有用于校准pH测量装置的参考材料可追溯到常见的pH标准,只能确保pH值的所需可比性。由OIM1 [3]目前最常使用的多主标准规模,并且由IUPAC [2]提供了许多固定点。这些固定点是国际商定标称组合物的缓冲溶液的pH值,使用没有液边的电化学电池测量。因此排除了扩散电位引起的不确定性。旨在提高欧洲pH值测量的均匀性,在1993/94年举行了欧洲初级pH测量程序的第一个比较欧洲项目300。该项目在参与实验室的pH结果之间表现出良好的一致性,除了在一种情况下,即邻苯二甲酸盐缓冲液[1]。为了继续这项工作,1996年进行了第二次Eutomet项目370“用于pH的标准测量装置的相互熟悉的pH值”[4]。本文描述了该项目,其中已经进行了四种缓冲材料的比较。再次测量邻苯二甲酸盐缓冲材料的pH。除了欧洲项目的合作伙伴之外,来自葡萄牙和英国的370个实验室加入了这一比较。葡萄牙和英国的合作伙伴没有记录的质量保证体系,因此他们不符合Eutomet项目的要求。这些参与者使用相同的测量程序和相同的材料,并且它们具有高精度pH测量的经验。出于这些原因,他们的结果包含在此处提供的报告中。所有合作伙伴都列于表1。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号