首页> 外文会议>National Physics Seminar >Design and development of a wireless medical device for blood leak detection during hemodialysis therapy
【24h】

Design and development of a wireless medical device for blood leak detection during hemodialysis therapy

机译:血液透析治疗期间血液泄漏检测无线医疗装置的设计与开发

获取原文

摘要

Hemodialysis therapy is a clinical intervention by inserting the needles on the arterial and venous blood vessel in the patient's arm. The area where the needle is inserted can cause bleeding and it's harmful for the patient. Previous method for detecting the blood vessel is by using a single point optical detection method. However, this method is not suitable for covering wide area. In this research, an electrochemical based multisensor is used for detecting the blood vessel around area of the inserted AV Fistula needles. This sensor is based on the impedance sensor. The main components of blood leak detection system are a multiplexer, an instrumentation amplifier, an impedance analyzer, microcontroller integrated Wi-Fi communication, and alarm system. An impedance analyzer is used to extract the signals into the real part and imaginary parts. The magnitude of the signals is computed from the real and imaginary parts of the signals. Such signals are computed using an artificial neural network (ANN) algorithm to classify the digitized signal output of the impedance analyzer whether the sample is human blood or not. The classification result in the form of messages is then sent to the alarm system to activate the visual and audio alarms. The messages were also transmitted to the nurse station through Wi-Fi communication system. Such message also consists of information about the location of the patient so the medical staf can go to the patient's bed directly. Various testing activities such functional test, battery life test, transmision test have been performed in this work. The result shows that the system can detect the present of the blood leakage with the sensitivity greater than 90%.
机译:血液透析治疗是通过在患者臂中插入动脉和静脉血管上的针来进行临床干预。插入针的区域会导致出血,对患者有害。以前用于检测血管的方法是通过使用单点光学检测方法。但是,这种方法不适合覆盖广域。在该研究中,使用电化学的多传感器用于检测插入的AV瘘针的区域周围的血管。该传感器基于阻抗传感器。血液泄漏检测系统的主要组件是多路复用器,仪表放大器,阻抗分析器,微控制器集成Wi-Fi通信和报警系统。阻抗分析器用于将信号提取到真实部件和虚部。信号的幅度从信号的实际和虚部计算。使用人工神经网络(ANN)算法计算这种信号,以对阻抗分析器的数字化信号输出进行分类,以是否存在样品是人类的血液。然后将消息形式的分类结果发送到警报系统以激活视觉和音频警报。通过Wi-Fi通信系统还将消息发送到护士站。这些消息还包括关于患者位置的信息,因此医疗STAF可以直接去患者的床。各种测试活动如此功能测试,电池寿命测试,传输测试已经在这项工作中进行。结果表明,该系统可以检测血液泄漏的目的,灵敏度大于90%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号