首页> 外文会议>SPIE Conference on Hard X-ray, Gamma-ray, and Neutron Detector Physics >Performance of an improved readout IC for multi-energy photoncounting x-ray detector arrays
【24h】

Performance of an improved readout IC for multi-energy photoncounting x-ray detector arrays

机译:用于多能量光子X射线探测器阵列的改进读出IC的性能

获取原文

摘要

We present the design and initial performance characterization of the XENA-2 readout IC for solid-state x-ray detector arrays. XENA-2 consists of 32 readout channels, each with charge-sensitive input amplifier, adjustable two-stage gain amplifier and five comparators with 16-bit pulse counters. Readout of the counters, over a 16-bit data bus, takes approximately 20 μs. Compared to the XENA chip, its predecessor, this new IC's main improvement is significantly reduced noise, which allows for lower comparator thresholds and increased count rates.
机译:我们介绍了固态X射线探测器阵列的Xena-2读出IC的设计和初始性能表征。 Xena-2由32个读数通道组成,每个读数通道,带充电输入放大器,可调节的两级增益放大器和具有16位脉冲计数器的五个比较器。在16位数据总线上读数计数器,大约为20μs。与Xena芯片相比,其前身,这一新的IC的主要改进显着降低了噪声,这允许更低的比较器阈值和增加的计数率。