首页> 外文会议>ASMS Conference on Mass Spectrometry and Allied Topics >Development of a compact multiple-ionization-stage TOF mass analyzer system for trace component monitoring within chemically challenging process gas matrices
【24h】

Development of a compact multiple-ionization-stage TOF mass analyzer system for trace component monitoring within chemically challenging process gas matrices

机译:在化学挑战过程气体矩阵内进行痕量组分监测的紧凑型多电离阶段TOF质量分析仪系统的研制

获取原文

摘要

Setup: A TOFWerk CTOF is successfully equipped with an ion notch filter and a custom ion transfer stage. The ion notch filter strongly suppresses selected m/z-ranges to improve the transmission efficiency of target analyte ions. The ionization chamber allows usage of several filament-free ionization methods, such as two-photon laser ionization, single-photon ionization, and chemical ionization. The latter is driven by a gas discharge, which operates at intermediate pressures of apporx. 1 mbar. With selected orifice sizes, sample pressures in the range 10~(-2) to 10~(-3) mbar are tolerated.
机译:设置:TOFWERK CTOF配备离子凹口滤波器和定制离子传输级。离子凹口滤波器强烈抑制所选择的M / Z范围,以提高目标分析物离子的传输效率。电离室允许使用几种无丝电离方法,例如双光子激光电离,单光子电离和化学电离。后者由气体放电驱动,该气体放电在Apporx的中压下运行。 1毫巴。采用所选择的孔尺寸,容忍10〜(-2)至10〜(-3)曼巴的样品压力。