Close-geometry efficiency calibration of p-type HPGe detectors with a Cs-134 point source

机译:具有CS-134点源的P型HPGE探测器的紧密几何效率校准

获取原文

摘要

When close-geometry detector calibration is required in gamma-ray spectrometry, single-line emitters are usually used in order toavoid true coincidence summing effects. We managed to overcome this limitation by developing a method for the determination of the efficiency of p-type HPGe detectors in close-geometry with a calibrated Cs-134 point source. No separate determination of coincidence summing correction factors is required and a single measurement furnishes the full-energy-peak efficiencies in the 475-1365 keV energy range.
机译:当在伽马射线光谱法中需要紧密几何探测器校准时,通常使用单线发射器,以便进行TAVOID真正的巧合求和效果。我们通过开发一种用于确定具有校准CS-134点源的紧密几何形状的p型HPGE检测器的效率来克服这种限制。不需要单独确定重合校正校正因子,单一测量提供475-1365 Kev能量范围内的全能峰效率。