VENTRICULAR PACING THRESHOLDS FOLLOWING HIGH-ENERGY VENTRICULAR DEFIBRILLATION SHOCKS

机译:高能量室外除颤冲击后心室起搏阈值

获取原文

摘要

Increased ventricular pacing thresholds have been observed following monophasic waveform shocks in implantable cardioverter defibrillators (ICDs). This study aimed to examine such changes following high-energy biphasic shocks in ICDs. Method: Ten episodes of VF were induced every 10 minutes in 10 pigs (23,l+-3.0 kg). After 10 seconds of VF a 40J biphasic shock (total 10 shocks) was delivered for successful defibrillation in the true-bipolar sensing lead system of the ICD. Ventricular bipolar pacing thresholds before and after these shocks were evaluated at one-minute intervals. Results: The mean pacing threshold before shock delivered was 0.066+-0.059 uJ. Those of the first,second and third minutes after the first shock were 0.052+-0.061 uJ,0.044+-0.039 irrespectively; showing that pacing thresholds gradually decreased. Conclusion: It may not be necessary to pace at a high-voltage output after biphasic shocks in ICDs.
机译:在可植入的心脏除颤器(ICDS)中单一的波形冲击之后,已经观察到室内花学阈值增加。本研究旨在在ICD中的高能量双相冲击后检查此类变化。方法:在10只猪(23,L + -3.0kg)中每10分钟诱导vf的十集。在VF的10秒之后,在ICD的真正双极传感铅系统中,提供了40J的双相休克(总10个冲击),以成功除颤。在每间隔一分钟评估这些冲击之前和之后的心室双极起搏阈值。结果:震动前的平均起搏阈值为0.066 + -0.059 UJ。第一次休克后的第一个,第二和第三分钟的那些是0.052±0.061 UJ,不断地0.044 + -0.039;显示起搏阈值逐渐减少。结论:在ICD的双相冲击后,可能没有必要在高压输出处于高压输出。