首页> 外文会议>International congress of imaging science >Reduction of patient dose on medical radiographs using scattered X-rays
【24h】

Reduction of patient dose on medical radiographs using scattered X-rays

机译:使用散射X射线减少医疗射线照片上的患者剂量

获取原文

摘要

We propose a new method for reduction of the patient dose by using scattered X-rays in order to achieve the same density as without scattered rays. The minimum perceptible thickness difference Δ_(min) of the object was calculated using psychophysical analysis for various radiographic densities, scatter fractions and luminous exitances of the viewer. The mAs values to obtain many densities were measured using four kinds of anti-scattered X-ray grid with their scatter fractions. The measured values were applied to above calculated psychophysical results to obtain the reduction ratio by using the scattered X-rays in actual radiography. The perceptibility of human eyes is influenced by luminous exitance of the viewer. As a result, the patient dose reduced 33 percent in maximum under the same perceptibility of Δ_(min) by increasing the luminous exitance from 1500 1m·m~(-2) to 8000 lm·m~(-2). The method of changing grid will be considered.
机译:我们提出了一种通过使用散射的X射线来减少患者剂量的新方法,以达到与没有散射的射线相同的密度。使用对观察者的各种射线照相密度,散射级分和发光出差的心理物理分析来计算物体的最小可感知厚度Δ_(min)。使用具有它们的散射级分的四种抗散射X射线栅格测量以获得许多密度的MAS值。将测量值施加到上面计算的心理物理结果,以通过在实际射线照相中使用散射的X射线来获得减少比率。人眼的可见性受到观察者的发光出口的影响。结果,通过增加15001M·m〜(-2)至8000lm·m〜(-2),患者剂量在相同的Δ_(min)的相同可知性下最大降低33%。将考虑改变网格的方法。