首页> 外文会议>Sensor Symposium on Sensors, Micromachines, and Applied Systems >Development and Evaluation of Microfluidic Device for Mn Ion Quantification in Ocean Environments
【24h】

Development and Evaluation of Microfluidic Device for Mn Ion Quantification in Ocean Environments

机译:海洋环境中Mn离子量化微流体装置的开发与评价

获取原文

摘要

This paper presents design, fabrication and evaluation of a polydimethylsiloxane (PDMS)-glass microfluidic device for quantitative determination of manganese ion (Mn(II)) in oceanic environments. The microfluidic device can perform luminol-hydrogen peroxide chemiluminescence (CL) reaction in a flow-through manner for highly sensitive Mn (II) quantification. A chelate resin column was integrated in the microfluidic device for removal of coexisting metal ion such as Fe(II). The developed device could quantify Mn(II) in nM level under coexistence of μM level of Fe(II). The microfluidic device developed here will be used as a core element of a totally integrated in situ chemical analyzer that we have been developing for oceanography applications.
机译:本文介绍了用于定量测定海洋环境中锰离子(Mn(II))的聚二甲基硅氧烷(PDMS)微流体装置的设计,制备和评价。微流体装置可以以流通方式对升氢过氧化氢化学发光(CL)反应用于高敏感的Mn(II)定量。将螯合树脂柱集成在微流体装置中,用于去除共存金属离子,例如Fe(II)。开发的装置可以在Fe(II)的共存下量化NM水平中的Mn(II)。这里开发的微流体装置将用作完全集成的核心元素,我们已经开发了海洋应用。