A Computerized Tomography Phantom: Design and Construction

机译:计算机化层析成像幻影:设计和施工

获取原文

摘要

A phantom for performance evaluation and quality assurance of CT scanners was designed and built in agreement with quality control requirements established by the Mexican Regulations (NOM-158-SSA1-1996). Phantom materials were chosen after experimental determination of their Hounsfield units. The phantom allows to perform the following tests: calibration, constancy and uniformity of the CT number, CT number dependence on the reconstruction algorithm, high and low contrast resolution, slice thickness and coincidence of the slice with the light system for patient alignment. We present and discuss the results of these test applied to two CT scanners.
机译:CT扫描仪的绩效评估和质量保证的幻影是由墨西哥法规建立的质量控制要求(NOM-158-SSA1-1996)的协议。在实验确定其Hounsfield单位后选择幻影材料。幻像允许执行以下测试:CT号的校准,恒定和均匀性,对重建算法的CT编号依赖性,具有用于患者对准的光系统的高和低对比度分辨率,切片厚度和切片的重合。我们展示并讨论这些测试的结果应用于两个CT扫描仪。