Calibration and Correction Factors of a Set of Semiconductor Diodes for its Use in External Radiation Therapy

机译:一组半导体二极管的校准和校正因子用于外部放射治疗的应用

获取原文

摘要

The use of semiconductor diodes at the Instituto Nacional de Ciencias Medicas y Nutricion Salvador Zubiran has been essential for in-vivo dose verification of treatments in Radiation Therapy. Three types of semiconductor diodes, two for photons of low (6 and 10 MV) and high (15 MV) energies; one for electron beams were characterized. The entrance and exit calibration factors for photons and the entrance calibration factors for electrons were obtained under standard conditions in a phantom of solid water using cylindrical and plane parallel ionization chambers for photons and electrons, respectively. Also, the signal diode as a function of monitor units and the angular dependence for photon and electron beams were analyzed; and the source surface distance and field size dependences for photons were determined. The results show that the signal is linear with monitor units and depends on beam quality, increases as a function of angle of incidence and field size and decreases with source surface distance.
机译:在Instituto Nacional de CienciaS Medicas Y Natricion萨尔瓦多Zubiran的使用在辐射治疗中的治疗中至关重要。三种类型的半导体二极管,两个用于低(6和10 mV)和高(15 mV)能量的光子;特征是电子束的一个。光子的入口和出口校准因子和电子的入口校准因子在使用用于光子和电子的圆柱形和平面的平行电离室的固体水体系的标准条件下获得。而且,分析了作为监测单元的函数的信号二极管和光子和电子束的角度依赖性;确定光子的源表面距离和场尺寸依赖性。结果表明,信号与监视器单元线性线性,取决于光束质量,随着入射角和场尺寸的函数而增加,随着源表面距离而减小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号