Commissioning of a Relative Stopping Power to Hounsfield Unit Calibration Curve for a Mevion Proton Radiation Treatment Unit

机译:用于羊膜质子辐射处理单元的Hounsfield单位校准曲线的相对停止动力

获取原文

摘要

A calibration curve relating relative stopping power (RSP) to Hounsfield units (HU) for proton treatment planning was commissioned. Synthetic tissue equivalent materials were imaged with varied scan parameters on two separate CT units to account for potential variations in Hounsfield units. All the materials were measured for their relative stopping power (RSP) with different energy proton beams from a Mevion passive double scattering proton unit, and the results were compared to Bethe-Block calculated values. The calculated relative stopping power (RSP) was found to match measured data within 2.5% in all cases except for low density lung. Hounsfield units (HU) were found to vary with scan parameters by less than 2.14% for near-tissue materials, and up to 7.44% for lung and bone. HU values between CT scanners were found to be comparable.
机译:将相对停止功率(RSP)与Hounsfield Units(HU)相关的校准曲线进行委托进行委托。在两个单独的CT单元上以各种扫描参数对合成组织等同材料进行成像,以考虑Hounsfield单元的潜在变化。针对它们的相对停止功率(RSP)测量所有材料,其具有来自百强无源双散射质子单元的不同能量质子束,并将结果与​​贝尔嵌段计算值进行比较。除了低密度肺的所有情况下,发现计算的相对停止功率(RSP)将测量数据匹配2.5%。发现Hounsfield Units(Hu)随扫描参数的扫描参数而异,肺部和骨骼的扫描参数不到2.14%,高达7.44%。 CT扫描仪之间的HU值被发现是可比的。