首页> 外文会议>Mexican Symposium on Medical Physics >Image Quality and Mean Glandular Dose in Mexican Mammography Services Measured with an Object-Insert/TL Mammography Kit
【24h】

Image Quality and Mean Glandular Dose in Mexican Mammography Services Measured with an Object-Insert/TL Mammography Kit

机译:用物体插入/ TL乳房X线摄影套件测量的墨西哥乳房X线摄影服务中的图像质量和平均腺剂量

获取原文

摘要

Thirty mammography systems were evaluated using an Object-Insert/TL kit (composed of an image quality and a dosimetric phantom) to assess image quality and mean glandular dose. All systems complied with the minimum score for object visibility in the image phantom, but most of the images showed artifacts. The average of the mean glandular dose was 1.7±0.7 mGy, and three systems delivered doses higher than the 3 mGy reference value. A weak correlation was found between image quality and dose, possibly reflecting the lack of quality control in the evaluated equipment.
机译:使用物体插入/ TL套件(由图像质量和剂量谱系组成)评估30个乳房X线摄影系统,以评估图像质量和平均腺剂量。所有系统均符合图像幻影对象可见性的最低分数,但大多数图像都显示了伪影。平均腺体剂量的平均值为1.7±0.7米,并且三种系统递送了高于3 MGY参考值的剂量。在图像质量和剂量之间发现了弱相关性,可能反映了评估设备中缺乏质量控制。