摘要

Standards and Calibration Laboratory of Hong Kong SAR, China has set up calibration service for spectral radiance light sources in the wavelength range from 300 nm to 1800 nm. The calibration is performed by direct comparison with reference spectral radiance light sources. A series of evaluation, including the study of the size of source effect, were conducted to validate the performance of the calibration system. In this paper, the evaluation of the size of source effect of the system is presented. The size of source effect was estimated as an uncertainty component of less than 0.1%. This is a relatively small contribution to the overall expanded measurement uncertainty of the system, which was estimated as 2.4% for the wavelength range between 400 nm to 1000 nm.
机译:香港特区的标准和校准实验室,中国已在波长范围为300nm至1800nm的波长范围内设置校准服务。通过与参考光谱光源光源直接比较来执行校准。进行了一系列评估,包括研究源效应大小的研究,以验证校准系统的性能。本文介绍了系统源效应大小的评价。源效应的尺寸估计为不确定性分量小于0.1%。这是对系统整体扩展测量不确定性的贡献相对较小,其估计为400nm至1000nm之间的波长范围的2.4%。