首页> 外文会议>International Symposium on Medical Measurements and Applications >Assessment of cardio-respiratory rates by non-invasive measurement methods in hospitalized preterm neonates
【24h】

Assessment of cardio-respiratory rates by non-invasive measurement methods in hospitalized preterm neonates

机译:通过无创测量方法评估住院早产儿的心脏呼吸频率

获取原文

摘要

Noncontact monitoring of physiological signsin preterm infants is necessary in order to increase neonatal comfort and open new possibilities to home respiration monitoring. The presented solution is obtained by using commercially available digital webcamera (WeC) and Eulerian video magnification (EVM) algorithm which amplify the small changes on the sequence of the recorded images, that would be invisible to the naked eye. With this method, the respiration rate (RR) is extracted as the dominant frequency in the thorax portion of the patient, and the heart rate (HR) by observing minor color changes of the skin. The results obtained are compared with the signal from standard multi-parameter monitor (MM). Data collected on 7 patients demonstrate the feasibility of the proposedmeasurement method. The measure of RR and HR show close correlation with the data from the MM, with a root mean square error of 12.2 for the HR and 7.6 for the RR. This technique can be a valuable non-contact monitoring system for the improvement of clinical care in NICU and may representa potential application in home health monitoring.
机译:为了增加新生儿的舒适度并为家庭呼吸监测开辟新的可能性,早产儿生理信号的非接触式监测是必要的。通过使用可商购的数字网络摄像头(WeC)和欧拉视频放大(EVM)算法获得所提出的解决方案,该算法放大了记录图像序列上的细微变化,而肉眼看不到这些细微变化。使用这种方法,通过观察皮肤的细微颜色变化,将呼吸频率(RR)提取为患者胸部的主要频率,并提取心率(HR)。将获得的结果与来自标准多参数监视器(MM)的信号进行比较。收集到的7位患者的数据证明了该测量方法的可行性。 RR和HR的度量与MM数据密切相关,HR的均方根误差为12.2,RR的均方根误差为7.6。该技术可能是用于改善重症监护病房临床护理的有价值的非接触式监测系统,并且可能代表了家庭健康监测中的潜在应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号