Relative reflectivity uncertainty evaluation for a broadband spectrophotometer system

机译:宽带分光光度计系统的相对反射率不确定度评估

获取原文

摘要

This study aims to analyze the reflectivity measurement uncertainty of a spectrophotometer, with considerations of the perpendicular incidence/reflection, scattering, and artificial manipulation. The operation procedure is designed to utilize a double beam spectrophotometric system, taking advantage of the direct comparison of the luminous intensity to evaluate the uncertainty. The environmental range of the temperature and the relative humidity for the measuring process are controlled to be 23±1.5oC and 50%±10%, respectively. The range for correction reflectivity is 10%-100%. High-reflectivity standard parts are used with the integrating sphere for the calibration. Low-reflectivity standard parts are used along with an integrating sphere for the evaluation when the measurement reflectivity is 0%-9%. The optimal expanded composition uncertainty of the system is found to be 1.2% in the wavelength range of 400 nm-700 nm.
机译:这项研究旨在分析分光光度计的反射率测量不确定度,并考虑垂直入射/反射,散射和人工操作。该操作程序设计为利用双光束分光光度系统,并利用直接比较发光强度来评估不确定度的优点。测量过程的温度和相对湿度的环境范围分别控制为23±1.5oC和50%±10%。校正反射率的范围是10%-100%。高反射率标准零件与积分球一起用于校准。当测量反射率为0%-9%时,将低反射率标准零件与积分球一起使用进行评估。发现在400 nm-700 nm的波长范围内,系统的最佳扩展成分不确定度为1.2%。