首页> 外文会议>Conference on physics of medical imaging >Evaluation of adaptation strengths of CARE Dose 4D in pediatric CT
【24h】

Evaluation of adaptation strengths of CARE Dose 4D in pediatric CT

机译:评价CARE剂量4D在儿科CT中的适应强度

获取原文

摘要

The motivation of this study is the general lack of knowledge regarding the efficiency and the appropriate use of the adaptation strengths of Siemens automatic exposure control system CARE Dose 4D. The purpose was to evaluate the effect on radiation absorbed dose using different adaptation strengths of CARE Dose 4D in three routine pediatric CT protocols. A pediatric anthropomorphic whole body phantom was used to simulate a 4 year old patient. CT scans were performed with a Siemens SOMATOM Definition Flash using three different pediatric protocols: neck, thorax, and abdomen. The characteristic of the tube current modulation was similar for all adaptation strengths. The difference is the extent of decrease in tube current. The degree of dose reduction using CARE Dose 4D and CARE kV compared using a fix effective mAs was 34-57%, 51-88%, and 56-91% for neck, thorax, and abdomen protocol, respectively. Accordingly, there is a large difference in radiation dose dependent on the adaptation strength: a factor of 1.5, 4.5, and 4.6 for neck, thorax, and abdomen protocol, respectively. The adaptation strengths can be used to obtain user-specified modifications of image quality or radiation dose to the patient. Radiologists and medical physicists need to be aware of the large differences between the adaptation strengths, and such differences are useful when attempting strategies to optimize CT radiation dose.
机译:这项研究的动机是,普遍缺乏有关西门子自动曝光控制系统CARE Dose 4D的效率和适应强度的适当使用的知识。目的是在三种常规的儿科CT方案中使用不同的CARE Dose 4D适应强度来评估对辐射吸收剂量的影响。儿科拟人化的全身模型被用来模拟一个4岁的病人。 CT扫描是使用Siemens SOMATOM Definition Flash使用三种不同的儿科规程进行的:颈部,胸部和腹部。对于所有适应强度,管电流调制的特性均相似。区别在于管电流减小的程度。与固定有效mAs相比,使用CARE Dose 4D和CARE kV减少剂量的颈部,胸部和腹部方案的减少程度分别为34-57%,51-8%和56-91%。因此,取决于适应强度的辐射剂量存在很大差异:分别针对颈部,胸部和腹部方案的系数分别为1.5、4.5和4.6。适应强度可用于获得用户指定的图像质量或对患者的辐射剂量的修改。放射科医生和医学物理学家需要意识到适应强度之间的巨大差异,这种差异在尝试优化CT辐射剂量的策略时非常有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号