首页> 外文会议>World Congress on Medical Physics and Biomedical Engineering >Comparison of the Three Filter Algorithms for Detection of Electrically-Evoked Short-Latency Responses in Retinal Ganglion Cells.
【24h】

Comparison of the Three Filter Algorithms for Detection of Electrically-Evoked Short-Latency Responses in Retinal Ganglion Cells.

机译:三种滤波算法的比较检测视网膜神经节细胞中电诱发短延迟应答的影响。

获取原文

摘要

When retinal ganglion cells are stimulated ex vivo by electrical current, short latency responses (within 30 ms) and long latency responses (over 30 ms) are evoked. The short-latency response is suspected to be originated from the direct electrical stimulation. For optimal stimulation protocol of retinal prosthesis, this short-latency response is getting important more and more. In this paper, we compared and evaluated performance of the three algorithms for the short-latency spike detection; suppression of artifacts by local polynomial approximation (SALPA), moving average filter (MAF), and forward-reverse filter (FR filter). The SALPA and the FR filter showed better detection performance comparing with the MAF algorithm.
机译:当通过电流刺激视网膜神经节细胞,唤起短期延迟响应(30ms内)和长延迟响应(超过30ms)。怀疑短期响应源自直接电刺激。对于视网膜假体的最佳刺激方案,这种短期延迟响应越来越重要。在本文中,我们对三个算法进行了比较和评估了三种算法的性能;通过局部多项式近似(SALPA),移动平均滤波器(MAF)和正向滤波器(FR滤波器)抑制伪影。 SALPA和FR滤波器显示出与MAF算法比较的更好的检测性能。