In-Vivo Validation of 3D Multi-Frequency Liver Shear Wave Absolute Vibro-Elastography with an xMATRIX Array

机译:3D多频肝脏剪切波绝对vibro-elastography的体内验证用Xmatrix阵列

获取原文

摘要

Liver fibrosis arises from chronic liver diseases, such as hepatitis B, C, and nonalcoholic steatohepatitis and can result in cirrhosis and death. Magnetic resonance elastography (MRE) is commonly regarded as the imaging-based gold-standard for fibrosis staging. With the aid of a state-of-the-art matrix array transducer, our previous 3D ultrasound shear wave absolute vibro-elastography (S-WAVE) imaging was able to generate hepatic stiffness measurements which are comparable to MRE. In this work, we introduce multi-frequency S-WAVE imaging with a matrix array transducer, to provide more robust and reliable measurements by shorter overall exam time. The system was characterized with three liver tissue phantoms of different elasticity using the MRE results as the ground truth. Six healthy volunteers and six patients who have chronic liver diseases were imaged. Our results indicate that measurements from multi-frequency 3D S-WAVE with a matrix array transducer correlated better to MRE, compared to the readings from the transient elastography method (FibroScan, Echosens).
机译:肝纤维化产生慢性肝病,如乙型肝炎,C和非酒精性脂肪肝炎,并且可以导致肝硬化和死亡。磁共振弹性显影(MRE)通常被认为是纤维化分期的成像金标准。借助于最先进的矩阵阵列换能器,我们之前的3D超声剪切波绝对振动弹性显影(S波)成像能够产生与MRE相当的肝刚度测量。在这项工作中,我们使用矩阵阵列换能器引入多频S波成像,以通过更短的整体考试时间提供更强大和可靠的测量。该系统的特征在于使用MRE结果作为地面真理的三种不同弹性的三种肝组织幽灵。六名健康志愿者和六名患有慢性肝病的患者进行了成像。我们的结果表明,与来自瞬态弹性摄影方法(Fibroscan,Shosens)的读数相比,来自具有矩阵阵列换能器的多频3D S波的测量与MRE更好地相关。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号