首页> 外文会议>IEEE International Nanoelectronics Conference >Relevance between the oro-nasal thermistor signal's derivative and the nasal pressure transducer signal in airflow monitoring
【24h】

Relevance between the oro-nasal thermistor signal's derivative and the nasal pressure transducer signal in airflow monitoring

机译:气流监测中口鼻热敏电阻信号的导数与鼻压力传感器信号之间的相关性

获取原文

摘要

Signals of an oro-nasal thermistor (Th) and a nasal air pressure transducer (NP) are widely used as surrogates of airflow for diagnosing sleep-disordered breathing, while the relationships between Th and NP were less discussed. To demonstrate the time derivative of Th is linear correlated with NP, a total of 15,053 epochs (defined 30-s as 1 epoch) clinical data were studied. The correlation coefficients between the time derivative of Th and NP in each epoch were calculated and its mean value is 0.87. It indicated that the time derivative of Th has a linear positive correlation with NP.
机译:口鼻热敏电阻(Th)和鼻气压传感器(NP)的信号被广泛用作诊断睡眠呼吸障碍的气流的替代方法,而Th和NP之间的关系则很少讨论。为了证明Th的时间导数与NP呈线性相关,共研究了15053个时代(定义为30个s为1个时代)。计算出每个时期Th和NP的时间导数之间的相关系数,其平均值为0.87。表明Th的时间导数与NP呈线性正相关。