首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Automatic Measurements of Choroidal Thickness in EDI-OCT Images
【24h】

Automatic Measurements of Choroidal Thickness in EDI-OCT Images

机译:EDI-OCT图像中脉络膜厚度的自动测量

获取原文

摘要

Enhanced Depth Imaging (EDI) optical coherence tomography (OCT) provides high-definition cross-sectional images of the choroid in vivo, and hence is used in many clinical studies. However, measurement of choroidal thickness depends on the manual labeling, which is tedious and subjective of interobserver differences. In this paper, we propose a fast and accurate algorithm that could measure the choroidal thickness automatically. The lower boundary of the choroid is detected by searching the biggest gradient value above the retinal pigment epithelium (RPE) and the upper boundary is formed by finding the shortest path of the graph formed by valley pixels using dynamic programming. The average of Dice’s Coefficient on 10 EDI-OCT images is 94.3%, which shows good consistency of the algorithm with the manual labeling. The processing time for each image is about 2 seconds.
机译:增强的深度成像(EDI)光学相干断层扫描(OCT)提供体内脉络膜的高清横截面图像,因此在许多临床研究中使用。然而,脉络膜厚度的测量取决于手动标签,这是乏味的乏味和interobserver差异的主观。在本文中,我们提出了一种快速准确的算法,可以自动测量脉络膜厚度。通过在视网膜颜料上高于视网膜颜料上高于(RPE)上方的最大梯度值并通过使用动态编程找到由谷像素形成的曲线图的最短路径来形成脉络膜的下边界。骰子系数的平均值为10 EDI-OCT图像为94.3%,其显示算法与手动标签的良好一致性。每个图像的处理时间约为2秒。