Can we use 1{sup left}H MRS shimming values to obtain 31{sup left}P spectra?

机译:我们可以使用1 {sup left} h夫人闪闪发光的值,以获得31 {sup left} p spectra?

获取原文

摘要

The perfect shimming of 31{sup left}P magnetic resonance spectroscopy (MRS) is not easy in vivo. The purpose of this study was to examine the feasibility of using 1{sup left}H MRS shimming values to obtain 31{sup left}P spectra in a same sequence. Both phantom and volunteer studies were carried out in this study. Phantom was a sphere filled with physiological metabolites of brain. In vivo study was performed on 4 healthy volunteers. The studies were performed on a 3-T GE scanner. A same localizer and a same cursor were used for both 1{sup left}H and 31{sup left}P scans. A spin echo MRS sequence was utilized for both 1{sup left}H scans with a standard head coil and 31{sup left}P scans with a GE service coil. 1{sup left}H scan was performed using first and automatic shimming and water linewidth (FWHM) of 3 Hz for phantom and 5 Hz for the volunteer were obtained. Shimming values of 1{sup left}H scan in x, y, z directions were copied to 31{sup left}P scan. A routine procedure of 31{sup left}P scan without value coping was also performed. Spectra were analyzed using SAGE/IDL program. Signal to noise ratio (SNR) was defined as the ratio of the signal height/maximum noise height. Good 1{sup left}H spectra and 31{sup left}P spectra were obtained for both phantom and volunteer studies. The 31{sup left}P spectra with 1{sup left}H MRS shimming values were similar with the 31{sup left}P spectra obtained with routine procedure. Lower SNRs of 31{sup left}P spectra were obtained in phantom with 1{sup left}H MRS shimming values, compared with routine procedure scan. Average SNR for Pi of 31{sup left}P spectra was 7.45:1 in the volunteer study with routine procedure, and 7.275:1 with 1{sup left}H MRS shimming values. 1{sup left}H MRS shimming values can be used to obtain useful 31{sup left}P spectra in a same sequence.
机译:完美的匀场31 {SUP左} P磁共振光谱(MRS)在体内并不容易。本研究的目的是检查使用1 {SUP左} H MRS Swimming值的可行性,以以相同的顺序获得31 {SUP左} P光谱。在本研究中进行了幻影和志愿者研究。幻影是一个充满脑的生理代谢物的球体。体内研究是对4个健康的志愿者进行的。这些研究是在3-T GE扫描仪上进行的。相同的定位器和相同的光标用于1 {sup left} h和31 {sup left} p扫描。利用标准头部线圈的1 {SUP左} H扫描和使用GE服务线圈扫描31°扫描,使用旋转回波MRS序列。 1 {SUP左} H扫描使用3 Hz的3 Hz的第一和自动遮光和水线宽(FWHM)进行幻影,并获得5 Hz的志愿者。在x,y,z方向上扫描1 {sup左} h扫描的屏幕值被复制到31 {sup left} p扫描。还执行了没有价值应对的31 {SUP左} P扫描的例程步骤。使用Sage / IDL程序分析光谱。信噪比(SNR)定义为信号高度/最大噪声高度的比率。良好的1 {SUP左} H光谱和31 {SUP左} P光谱是幻象和志愿者研究。带有1 {SUP左} H MRS Shimming值的31 {SUP左} P光谱与使用例行程序获得的31 {SUP左} P光谱相似。与常规过程扫描相比,在幻影中获得了31 {SUP左} P光谱的较低的SNR,与1 {SUP左} H MRS Shimming值。 PI的平均SNR为31 {SUP} P光谱为7.45:1,在志愿者学习中,常规程序,7.275:1,1 {SUP左} H MRS Shimming值。 1 {sup left} h夫人闪闪发光值可用于以相同的顺序获得有用的31 {sup左} p光谱。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号