Investigating the Tightening Torque of an Osseointegrated Transfemoral Implant System

机译:调查骨整合的经熔血液植入系统的紧固扭矩

获取原文

摘要

Osseointegrated transfemoral implants have been introduced as an alternative to conventional, socket-based prostheses for above knee amputees. This method is showing great promise; however, the frequency of device failure is of concern. As a means to understand some of these failures, this study investigates the manufacturer's specified tightening torque and the preload stress experienced by the retention bolt of the transfemoral system. This has been explored using a protocol proven during tests of dental implants with a similar assembly. Elongation of the bolt was recorded after applying known torques using a digital torque wrench, to a maximum of the manufacturer's specified torque. Using the elongation data, preload stress and force acting on the bolt were calculated. At the manufacturer's specified tightening torque, preload stress peaked at 23.8% of the material's yield strength, with the retention bolt tensile preload measured to be 3.5 kN. This suggests that the retention bolts may be tightened beyond the manufacturer's recommended tightening torque to increase assembly strength without inducing plastic deformation of the components.
机译:已经引入了骨整合的转发血液植入物作为膝关节上方的常规套接字的假体的替代方法。这种方法显示出很大的承诺;然而,设备故障的频率是有关的。作为一种理解其中一些故障的手段,本研究调查了制造商的特定拧紧扭矩和经熔体系统的保持螺栓所经历的预载力。这已经探讨了在具有类似组件的牙科植入物期间经过验证的协议探索。在使用数字扭矩扳手将已知的扭矩施加到最大的制造商的指定扭矩之后,记录螺栓伸长率。计算使用伸长率数据,计算出在螺栓上作用的预加载应力和力。在制造商的规定紧固扭矩下,预加载应力达到材料屈服强度的23.8%,保持螺栓拉伸预载荷为3.5kN。这表明可以在制造商推荐的紧固扭矩超出制造商的保留螺栓以增加组装强度而不会诱导组件的塑性变形。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号