摘要

Optical Coherence Tomography catheters comprise a transparent tube which can act as a negative cylindrical lens and introduce astigmatism which will lead to a decrease in transverse resolution and image contrast. In this report, we numerically analyzed the astigmatism for standard catheter designs applicable to esophageal and coronary imaging, hi order to maintain image quality, generally the beam can be refocused by a curved interface. To handle a situation involving high-index flush media, another method based on matching refractive indices is described and shown to successfully restore a round beam.
机译:光学相干断层扫描导管包括透明管,该透明管可以用作负圆柱透镜并引入像散,这将导致横向分辨率和图像对比度降低。在本报告中,我们对适用于食管和冠状动脉成像的标准导管设计的散光进行了数值分析。为了保持图像质量,通常可以通过弯曲界面重新聚焦光束。为了处理涉及高折射率冲洗介质的情况,描述并显示了基于匹配折射率的另一种方法,该方法成功地还原了圆光束。