首页> 外文期刊>Neuromodulation: journal of the International Neuromodulation Society >The Cross-Talk Between Spinal Cord Stimulators and the Confirm? Cardiac Monitor
【24h】

The Cross-Talk Between Spinal Cord Stimulators and the Confirm? Cardiac Monitor

机译:脊髓刺激器之间的对话和确认?心脏监护仪

获取原文
获取原文并翻译 | 示例
           

摘要

Introduction: Spinal cord stimulators (SCS) function by transmission of electrical impulses to electrode contacts placed within the epidural space depending on the painful area to be treated. Because of the electrical nature of the SCS, there has been concern about the interaction between these devices and devices that monitor or augment the cardiac system. Implantable loop recorders help to identify the causes of syncope or palpitation by continuously evaluating and recording portions of an electrocardiograph in patients being evaluated for cardiac conduction arrhythmias. The purpose of the study is to simulate the possible effects of spinal cord stimulation on a Confirm cardiac monitor (St. Jude Medical, St. Paul, MN, USA). Methods: Twenty patients without preexisting cardiac disease, who were successfully being treated with SCS, were enrolled. Confirm loop recorders (St. Jude Medical) were placed on their chest wall in a noninvasive manner, with all programmed at identical settings. Multiple stimulation settings were adjusted on the stimulators and the recordings from the Confirm loop recorder were analyzed for evidence of interference. Results: Fifteen of the patients had no electrical noise detected at any of the tested combinations of stimulation. Five patients had some electrical "noise" detected by the loop recorder, but it did not inhibit the cardiologist evaluating the recording from analyzing the electrocardiograph for diagnostic purposes. At no point with any of the patients at any tested setting was there an appearance of a life-threatening arrhythmia. Conclusion: Our study demonstrates that spinal cord stimulation is unlikely to interfere with the data collected by the Confirm loop recorder, and the presence of an SCS should not interfere with the ability to use a loop recorder for diagnostic purposes.
机译:简介:脊髓刺激器(SCS)通过将电脉冲传输到硬膜外腔内的电极触点来传输功能,具体取决于要治疗的疼痛区域。由于SCS的电性质,已经关注这些装置与监视或增强心脏系统的装置之间的相互作用。植入式环路记录仪可通过连续评估和记录正在评估心脏传导性心律不齐的患者的心电图仪部分,来帮助确定晕厥或心pa的原因。该研究的目的是在确认心脏监护仪(美国明尼苏达州圣保罗的圣裘德医疗公司)上模拟脊髓刺激的可能影响。方法:招募了20名成功接受SCS治疗的无既往心脏病的患者。确认循环记录仪(St. Jude Medical)以无创方式放置在胸壁上,所有程序均设置为相同设置。在刺激器上调整了多种刺激设置,并对来自Confirm Loop记录器的记录进行了分析以寻找干扰的证据。结果:在测试的任何刺激组合中,十五名患者均未检测到电噪声。五名患者的环路记录仪检测到一些电子“噪音”,但这并没有阻止心脏病专家评估记录以分析用于诊断目的的心电图仪。任何接受测试的患者在任何情况下都没有出现危及生命的心律不齐。结论:我们的研究表明,脊髓刺激不太可能干扰Confirm Loop记录仪收集的数据,并且SCS的存在不应影响使用Loop Recorder进行诊断的能力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号