Noninvasive blood pressure measurements are difficult when arterial pulsations are reduced, as in patients supported by continuous flow left ventricular assist devices (cf-LVAD). We evaluated the feasibility of measuring noninvasive arterial blood pressure with the Nexfin monitor during conditions of reduced arterial pulsatility. During cardiopulmonary bypass (CPB) in which a roller pump based or a centrifugal pump based heart-lung machine generated arterial blood pressure with low pulsatility, noninvasive arterial pressures (NAP) measured by the Nexfin Monitor were recorded and compared with invasively measured radial artery pressures (IAP). We also evaluated NAP in 10 patients with a cf-LVAD during a pump speed change procedure (PSCP). During CPB in 18 patients, the NAP-IAP average difference was -1.3 +/- 6.5 mm Hg. The amplitude of pressure oscillations were 4.3 +/- 3.8 mm Hg measured by IAP. Furthermore, in the cf-LVAD patients, increase in pump speed settings led to an increase in diastolic and mean arterial pressures (MAP) while the NAP acquired a sinusoidal shape as the aortic valve become permanently closed. In conclusion, NAP was similar to IAP under conditions of reduced arterial pulsatility. The device also measured the blood pressure waveform noninvasively in patients supported by a cf-LVAD. ASAIO Journal 2010; 56:221-227
机译:当减少动脉搏动时,非侵入性血压测量将很困难,例如在连续流左心室辅助设备(cf-LVAD)支持的患者中。我们评估了在动脉搏动降低的情况下,使用Nexfin监护仪测量无创性动脉血压的可行性。在心肺旁路(CPB)期间,基于滚子泵或基于离心泵的心肺机产生低脉动的动脉血压,记录由Nexfin Monitor测量的非侵入性动脉压(NAP)并将其与侵入性测量的radial动脉压进行比较(IAP)。我们还在泵速更改程序(PSCP)期间评估了10例cf-LVAD患者的NAP。在18例患者进行CPB期间,NAP-IAP平均差异为-1.3 +/- 6.5 mm Hg。通过IAP测得的压力振荡幅度为4.3 +/- 3.8 mm Hg。此外,在cf-LVAD患者中,泵速设置的增加导致舒张压和平均动脉压(MAP)的增加,而当主动脉瓣膜永久关闭时,NAP呈正弦曲线形状。总之,在动脉搏动降低的情况下,NAP与IAP相似。该设备还无创地测量了由cf-LVAD支持的患者的血压波形。 《 ASAIO杂志》 2010年; 56:221-227

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号