Assessment of the head-stem taper junction requires the estimation of material loss from the taper surfaces of both femoral head and stem. This paper describes a method for the measurement and analysis of material loss from the modular taper junction of hip replacements, in particular femoral stem tapers where generally the entire taper surface has been engaged. In such cases no direct unworn datum is readily identifiable to assess material loss. The highly anisotropic topology of some stem designs poses additional challenges to the measurement and analysis process. Estimation of material loss of retrieved femoral stems is further complicated by retrieval damage or surface deposits often present on the taper surface. The femoral head tapers typically exhibit areas of pristine surface attributed to the difference in taper length compared to the engaging stem. These areas can be selected as unworn when employed in the analysis process, provided they do not show surface damage or deposits. Measurement of the taper surfaces has been performed using a Talyrond (Ametek, Inc., US) out-of-roundness measurement instrument equipped with a 5µm diamond tip stylus. Vertical axial traces were employed to digitize the surface of the taper. Measurement data has been analyzed using a multi stage process that has been specifically adapted for stem tapers. The underlying stem taper geometry is determined by means of a morphological filter to remove the high aspect ratio microstructure. This paper presents a study of 40 retrieved LHMoM hip replacements that have been analyzed to ascertain the material loss at the modular taper junction. The purpose of this study was to ascertain the viability of characterizing material loss from the stem taper junction and to provide insight into the overall material loss contribution.
机译:评估头干锥形连接需要从股骨头和茎的锥形表面估计材料损失。本文介绍了一种用于测量和分析髋关节置换术(尤其是通常已接合整个锥面)的股骨柄锥度的模块化锥度连接处的材料损失的方法。在这种情况下,无法直接确定未磨损的原点以评估材料损失。一些阀杆设计的高度各向异性拓扑对测量和分析过程提出了额外的挑战。取回的股骨柄的材料损失的估计由于取回损伤或通常在锥形表面上存在的表面沉积而进一步复杂化。股骨头锥度通常表现出原始表面的面积,这归因于与接合杆相比锥度长度的差异。这些区域在分析过程中使用时可以选择为未磨损,前提是它们不显示表面损坏或沉积。锥形表面的测量是使用配有5µm金刚石尖端测针的Talyrond(美国Ametek,Inc.)不圆度测量仪进行的。垂直轴向迹线被用来数字化锥度的表面。使用专门针对茎锥度的多阶段过程分析了测量数据。底层的锥度几何形状通过形态过滤器确定,以去除高深宽比的微结构。本文介绍了一项对40例LHMoM髋关节置换术的研究,已对其进行了分析,以确定在模块化锥度连接处的材料损失。本研究的目的是确定表征茎锥连接处材料损失的可行性,并提供对总体材料损失贡献的了解。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号