A 46-year-old man presented to our institution with inappropriate implantable cardioverter-defibrillator (ICD) shock delivery. The ICD (single chamber, dual shock coils) was implanted for sustained monomorphic ventricular tachycardia with unstable hemodynamics and underlying systolic left ventricular dysfunction. ICD interrogation revealed recurrent episodes of ICD shock due to noise sensing and increased impedance of right ventricular-lead. With the impression of lead fracture, ICD lead extraction was performed. The fractured ICD lead was completely removed by traction of locking stylet and counter-traction of polypropylene dilator sheath. A new lead was inserted and the patient was discharged without complications after 2 days. To our knowledge, this is the first report on ICD lead extraction by conventional traction and counter-traction technique in Korea.
机译:一名46岁的男子因不适当的植入式心脏复律除颤器(ICD)电击送至我们机构。植入ICD(单腔,双电击线圈)可用于持续的单形性室速,具有不稳定的血流动力学和潜在的收缩期左心室功能障碍。 ICD询问显示,由于噪声感应和右心室导线阻抗增加,ICD休克复发。带着铅断裂的印象,进行了ICD铅提取。通过锁定管心针的牵引和聚丙烯扩张器护套的反向牵引,可以完全去除断裂的ICD引线。插入新导线,2天后患者无并发症出院。据我们所知,这是韩国首次通过传统牵引和反牵引技术提取ICD铅的报告。