近赤外分光画像計測法による無侵襲抹消血管モニタリング装置を用いたヘモグロビン濃度の測定

机译:使用无创性外周血管监测装置通过近红外光谱图像测量方法测量血红蛋白浓度

摘要

Measurement of hemoglobin (Hb) concentrations was performed using the noninvasive peripheral blood vessel monitoring device with the near-infrared spectroscopic imaging method. The percent coefficient of variation (CV) for the within-run precision of this method ranged from 0.85 to 8.1%. The CV for the within-day variation of this method ranged from 2.66 to 8.38%. When compared this method with the conventional automated analyzer of blood cells, these two assay methods were significantly correlated. The results of this study suggest that this method could be a useful screening test for Hb concentrations. It could be widely used as not only medical application but application in a healthy care field.
机译:血红蛋白(Hb)浓度的测量是使用无创性外周血管监测设备和近红外光谱成像方法进行的。该方法的内部运行精度的变异百分数(CV)为0.85至8.1%。此方法的日内变化的CV在2.66至8.38%之间。当将该方法与常规的血细胞自动分析仪进行比较时,这两种测定方法显着相关。这项研究的结果表明,该方法可以作为对Hb浓度的有用筛选测试。它不仅可以广泛地用作医学应用,而且可以广泛应用于健康护理领域。