摘要

Ab-initio study on modification of commerce terahertz spectrometer with time resolution Z-3 (Zomega, USA) by substitution of ZnTe and GaP detectors and LT-GaAs generator for homemade of pure and S-doped GaSe is carried out. It was established that in spite of not optimized parameters pure and doped GaSe:S(0.3 mass%) crystal are comparable, relatively, in generation efficiency and detection sensitivity to commerce units due to lower nonlinear optical loss and much higher damage threshold. The advantages are in force from pump fluences of below 5 mJ/cm{2} for pure GaSe. The closer S-doping to optimal concentration, the lover fluences resulting in the advantages. Pure and S-doped GaSe demonstrate higher reliability and larger dynamic range of operation. Recorded absorption spectra well match known spectra.
机译:进行了从头开始研究用时间分辨率为Z-3的商用太赫兹光谱仪(美国Zomega)替代ZnTe和GaP检测器以及LT-GaAs发生器以制备纯净和S掺杂的GaSe的修改。已经确定,尽管没有优化参数,但由于较低的非线性光学损耗和较高的损伤阈值,纯的和掺杂的GaSe:S(0.3质量%)晶体在生成效率和对商业单位的检测灵敏度方面具有相对的可比性。对于纯GaSe,低于5 mJ / cm {2}的泵注量是有好处的。 S掺杂越接近最佳浓度,情人就可以产生优势。纯和S掺杂的GaSe表现出更高的可靠性和更大的动态工作范围。记录的吸收光谱与已知光谱完全匹配。