首页> 美国政府科技报告 >Use of a Non-Linear Method for Including the Mass Uncertainty of Gravimetric Standards and System Measurement Errors in the Fitting of Calibration Curves for XRFA Freeze-Dried UNO sub 3 Standards
【24h】

Use of a Non-Linear Method for Including the Mass Uncertainty of Gravimetric Standards and System Measurement Errors in the Fitting of Calibration Curves for XRFA Freeze-Dried UNO sub 3 Standards

机译:使用非线性方法计算重量标准的质量不确定度和XRFa冷冻干燥的UNO sub 3标准校准曲线拟合中的系统测量误差

获取原文

摘要

A sophisticated nonlinear multiparameter fitting program was used to produce a best fit calibration curve for the response of an x-ray fluorescence analyzer to uranium nitrate, freeze dried, 0.2% accurate, gravimetric standards. The program is based on unconstrained minimization subroutine, VA02A. The program considers the mass values of the gravimetric standards as parameters to be fit along with the normal calibration curve parameters. The fitting procedure weights with the system errors and the mass errors in a consistent way. The resulting best fit calibration curve parameters reflect the fact that the masses of the standard samples are measured quantities with a known error. Error estimates for the calibration curve parameters can be obtained from the curvature of the ''Chi-Squared Matrix'' or from error relaxation techniques. It was shown that nondispersive XRFA of 0.1 to 1 mg freeze-dried UNO sub 3 can have an accuracy of 0.2% in 1000 s. (ERA citation 03:048427)