首页> 美国政府科技报告 >Gabor zone plate apertures for imaging with the mercuric iodide gamma ray camera.
【24h】

Gabor zone plate apertures for imaging with the mercuric iodide gamma ray camera.

机译:Gabor区域板孔用于使用碘化汞伽马射线照相机进行成像。

获取原文

摘要

Gabor zone plate (GZP) apertures have been developed for use in EG&G EM's mercuric iodide (H(sub g)I(sub 2)) gamma-ray camera. The purpose of such an aperture is to increase efficiency, while maintaining good resolution. The GZP is similar to the Fresnel zone plate (FZP) but it has continuous transitions between opaque and transparent regions. Because there are no sharp transitions in the transmission the inherent interference noise in GZP imaging is lower than that in FZP imaging. GZP parameters were chosen by considering the effects of constraints such as detector pixel size, number of pixels, minimum field of view required, maximum angle of incidence tolerated, and the Nyquist criterion for the minimum sampling rate. As a result an aperture was designed and fabricated with eight zones and a diameter of 3 cm. Lead was chosen as the aperture medium due to its high attenuation coefficient. Experimental data were obtained from the camera with the above GZP aperture. The point-spread function was determined and compared to the calculated response. Excellent agreement was obtained. The reconstruction process involves simulating, by computer, planar-wave illumination of a scaled transparency of the image and recording the intensity pattern at the focal plane. 6 refs., 7 figs. (ERA citation 15:032973)

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信