首页> 美国政府科技报告 >Structural Analysis and Optimization of the Support Device Used for a Proximal Fracture of the Femur
【24h】

Structural Analysis and Optimization of the Support Device Used for a Proximal Fracture of the Femur

机译:股骨近端骨折支撑装置的结构分析与优化

获取原文

摘要

The support system for a proximal fracture of the femur is studied by means of the finite element method. The support system of the Gamma III by Stryker is used for this study due to the versatility and simplicity of the system. The variance of each system was modeled to validate the results. Parametric studies between the models were conducted and design improvements were then tested against the existing parameters. Given the material properties for the Ti-6 AL-4V support system and bone, the models were considered under two different load conditions. From these results, a recommendation can be made to the manufacturing and medical community as to the effective use of the Gamma III nail implant system for a proximal fracture in the femur.