首页> 美国政府科技报告 >Calibration of Radiosondes at Temperatures to Minus 100 Degrees C
【24h】

Calibration of Radiosondes at Temperatures to Minus 100 Degrees C

机译:在不到100摄氏度的温度下校准无线电探空仪

获取原文

摘要

A method and apparatus to calibrate the thermistors of radiosondes are described. The method is based on cooling alcohol and toluene to -95 to -100C and a method for checking the readings of test thermometers at these temperatures against reference points. The method utilizes the familiar relationship between the sublimation point of dry ice and the external pressure.