Air-kerma strength determination of a 169Yb high dose rate brachytherapy source.

机译:169Yb高剂量率近距离放射治疗源的空气比释动能强度测定。

获取原文
获取原文并翻译 | 示例
           

摘要

The increased demand for high dose rate (HDR) brachytherapy as an alternative to external beam radiotherapy has led to the introduction of a HDR brachytherapy isotope 169Yb. This source offers a dose rate similar to 192Ir HDR sources, at about one fourth the effective photon energy. This work presents the calibration of this source in terms of air-kerma strength, based on an adaptation of the current, National Institute of Standards and Technology traceable, in air measurement technique currently used for 192Ir HDR sources. Several additional measurement correction factors were required, including corrections for air scatter, air attenuation, and ion recombination. A new method 169Yb is introduced for determining the ion chamber calibration coefficient Nk(169Yb). An uncertainty analysis was also performed, indicating an overall measurement expanded uncertainty in the air-kerma strength (k=2) of 2.2%.
机译:对于高剂量率(HDR)近距离放射疗法以替代外部束放射疗法的需求增加,导致引入了HDR近距离放射疗法同位素169Yb。该源提供的剂量率类似于192Ir HDR源,约为有效光子能量的四分之一。这项工作基于对当前可用于192Ir HDR源的空气测量技术的适应性,国家标准技术研究所的改编,提出了根据空气比释动能强度对该源进行的校准。还需要几个其他的测量校正因子,包括空气散射,空气衰减和离子重组的校正。引入了一种新的方法169Yb来确定离子室校准系数Nk(169Yb)。还进行了不确定性分析,表明总体测量将空气比释动能强度(k = 2)的不确定性扩大了2.2%。