Continuous non-invasive finger arterial pressure monitoring reflects intra-arterial pressure changes in children undergoing cardiac surgery.

机译:连续的无创手指动脉压监测反映了接受心脏手术的儿童的动脉内压变化。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Continuous non-invasive measurement of finger arterial pressure (FAP) is a reliable technology in adults. FAP is measured with an inflatable cuff around the finger and simultaneously converted to a reconstructed brachial artery pressure waveform (reBAP) by the Nexfin device. We assessed the adequacy of a prototype device (Nexfin-paediatric), designed for a paediatric population, for detecting rapid arterial pressure changes in children during cardiac surgery. METHODS: Thirteen anaesthetized children with a median age of 11 months (2 months-7 yr) undergoing congenital cardiac surgery were included in the study. reBAP and intra-arterial pressure (IAP) were recorded simultaneously during the surgical procedure. To assess the accuracy of reBAP in tracking arterial pressure changes, the four largest IAP variations within a 5 min time interval were identified from each procedure. These variations were compared offline with reBAP during a 10 s control period before and a 10 s period after an arterial pressure change had occurred. RESULTS: In 10 out of 13 children, a non-invasive arterial pressure recording could be obtained. Therefore, recordings from these 10 children were eligible for further analysis, resulting in 40 data points. The correlation coefficient between reBAP and IAP in tracking mean arterial pressure (MAP) changes was 0.98. reBAP followed changes in IAP with a mean bias for systolic, diastolic arterial pressure, and MAP of 0.0 mm Hg (sd 5.8), 0.1 (sd 2.8), and 0.19 (sd 2.7), respectively. CONCLUSIONS: The prototype device closely follows arterial pressure changes in children. However, in a considerable number of attempts, obtaining a signal was time-consuming or unsuccessful. This technique seems promising but requires further technical development.
机译:背景:连续无创测量手指动脉压(FAP)是成人的一项可靠技术。 FAP用手指周围的可充气袖套测量,并通过Nexfin设备同时转换为重建的肱动脉压力波形(reBAP)。我们评估了专为儿科人群设计的原型设备(Nexfin-儿科)的适当性,该设备可检测心脏手术期间儿童的快速动脉压变化。方法:13名接受中枢性心脏手术的麻醉儿童中位年龄为11个月(2个月至7岁)。在手术过程中同时记录了reBAP和动脉内压(IAP)。为了评估reBAP在跟踪动脉压变化中的准确性,从每个程序中识别出5分钟时间间隔内的四个最大IAP变化。在发生动脉压变化之前的10 s控制期间和之后的10 s期间,将这些变化与reBAP进行离线比较。结果:在13名儿童中,有10名获得了无创性动脉压记录。因此,这10名儿童的录音符合进一步分析的条件,得出40个数据点。 reBAP和IAP在跟踪平均动脉压(MAP)变化中的相关系数为0.98。 reBAP跟随IAP的变化,其收缩压,舒张动脉压和MAP的平均偏差分别为0.0 mm Hg(sd 5.8),0.1(sd 2.8)和0.19(sd 2.7)。结论:原型装置密切跟踪儿童的动脉压变化。然而,在大量尝试中,获得信号是费时的或不成功的。该技术似乎很有希望,但需要进一步的技术开发。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号