首页> 外文期刊>Bulletin du Bureau National de Metrologie >Comparaison COFRAC interlaboratoire de kerma dans 1'air et de dose absorbée dans I'eau pour les photons du cobalt 60
【24h】

Comparaison COFRAC interlaboratoire de kerma dans 1'air et de dose absorbée dans I'eau pour les photons du cobalt 60

机译:钴60光子的空气比释动能和水中吸收剂量的实验室间COFRAC比较

获取原文
获取原文并翻译 | 示例
           

摘要

In 1996, the standard laboratories accredited in dosimetry by COFRAC have been for the first time made directly traceable to the national standard of absorbed dose to water for 6 Co y-rays. For this radiation, their traceability was formerly established only in air kerma. The aim of this interlaboratory comparison, organised in the frame of COFRAC, was to test the coherence of the calibrations carried out by these laboratories for both quantities. It consisted in comparing, for a radiotherapy type dosimeter, the calibration coefficients obtained by the participants in their own 6 Co beams, to the values of BNM-LPRI. Results were in good agreement. Some problems arising from calibrations in absorbed dose to water, placing at the reference point the effective point of measurement of the ionization chamber instead of its centre, were pointed out. In that case, the differences in irradiation conditions during calibration and measurements (especially the source-to-detector distance,) can result in errors of some tenths of percent.
机译:1996年,首次通过COFRAC进行剂量学认证的标准实验室可直接追溯到6 Co y射线对水吸收剂量的国家标准。对于这种辐射,以前只能在空气比释动能中确定其可追溯性。在COFRAC框架内进行的实验室间比较的目的是测试这些实验室对这两个数量进行的校准的一致性。对于放疗型剂量计,它包括将参与者在其自己的6个Co光束中获得的校准系数与BNM-LPRI的值进行比较。结果非常吻合。指出了校准水吸收剂量所引起的一些问题,这些问题是将电离室的有效测量点置于参考点而不是其中心。在那种情况下,在校准和测量过程中辐照条件的差异(尤其是光源到探测器的距离)可能导致百分之十分的误差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号