首页> 外文期刊>Investigative radiology >Ultrasound-guided biopsy of focal lesions using three-dimensional ultrasound with a matrix array transducer: comparison with 2-dimensional ultrasound in a phantom study.
【24h】

Ultrasound-guided biopsy of focal lesions using three-dimensional ultrasound with a matrix array transducer: comparison with 2-dimensional ultrasound in a phantom study.

机译:使用三维超声与矩阵阵列换能器对局灶性病变进行超声引导活检:在幻像研究中与二维超声进行比较。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: To investigate the feasibility of biopsy-guided 3-dimensional (3D) ultrasound (US) with a matrix array transducer (3D US-MAT), using the lateral and/or elevation tilt modes. MATERIALS AND METHODS: Both inexperienced and experienced examiners performed biopsy experiments using a freehand technique on agar-based phantoms with 2-dimensional (2D) US guidance and 3D US-MAT guidance allowing real-time biplane imaging, using the lateral tilt mode and/or elevation tilt mode. Each phantom contained 4 targets of different sizes, and there were 3 different types of phantom, each having a different distance between the target and the surface. In every session, each target was aimed at 3 times, with 3 different distances between the transducer and puncture site, which resulted in 36 biopsies (3 biopsies per target x 4 targets x 3 phantoms, each of which was uniquely identified by size, depth, and angle). This procedure was carried out by each of 2 examiner twice by 2D US and twice by 3D US-MAT, resulting in paired data for each biopsy and each examiner. Target variables were target specimen length (TSL), procedure time, and agreement between the TSL of the first and second procedure by each examiner with identical biopsy parameters as a reliability measure. RESULTS: The intraexaminer agreements between the TSLs with the 3D US-MAT in both inexperienced (r = 0.84) and experienced (r = 0.93) examiners were higher than with the 2D US (r = 0.59 with P = 0.0066, and r = 0.87 with P = 0.06, respectively). However, the procedure time with the 3D US-MAT was significantly longer than with the 2D US. The TSL were significantly longer with 3D US-MAT than with 2D US for both the examiners; the inexperienced examiner (P < 0.001) benefited more than his experienced colleague (P = 0.024). In addition, the 3D US-MAT had significant benefits for the acquisition of TSL with small target diameters and for targets with shallow locations; with more acute puncture angles for both the examiners (all P < 0.05). CONCLUSIONS: Using a 3D US-MAT guidance improves the reliability and precision of biopsies, particularly under difficult conditions and for inexperienced examiners.
机译:目的:使用横向和/或仰角倾斜模式,研究使用矩阵阵列换能器(3D US-MAT)进行活检引导的3维(3D)超声(US)的可行性。材料和方法:没有经验的和经验丰富的检查员均使用徒手技术对琼脂模型进行活检实验,该假体具有二维(2D)US引导和3D US-MAT引导,可使用侧向倾斜模式和/或实时双平面成像或仰角倾斜模式。每个体模包含4个大小不同的目标,并且存在3种不同类型的体模,每个体模与表面之间的距离都不同。在每个阶段中,每个目标都瞄准了3次,换能器与穿刺点之间的距离为3,这导致进行了36次活检(每个目标3次活检x 4个目标x 3个幻影,每个都通过大小,深度来唯一标识和角度)。由2位检查员分别通过2D US和3D US-MAT进行两次此程序,从而得出每个活检和每个检查员的配对数据。目标变量是目标样本长度(TSL),操作时间以及每个检查员在第一和第二次操作的TSL之间的一致性,并以相同的活检参数作为可靠性测量。结果:没有经验的(r = 0.84)和有经验的(r = 0.93)的TSL与3D US-MAT的审查员内部协议高于2D US(r = 0.59,P = 0.0066,r = 0.87)分别为P = 0.06)。但是,使用3D US-MAT的过程时间明显长于使用2D US的过程时间。两位审查员使用3D US-MAT的TSL明显长于使用2D US的TSL。没有经验的考官(P <0.001)比他经验丰富的同事(P = 0.024)受益更多。此外,3D US-MAT对于获得目标直径较小的TSL和位置较浅的目标都具有显着优势。两位检查者的穿刺角度都更大(所有P <0.05)。结论:使用3D US-MAT指南可提高活检的可靠性和精确度,尤其是在困难的条件下以及对于没有经验的检查员而言。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号