The need for viral load (VL) monitoring of HIV patients receiving antiretroviral therapy (ART) in resource-limited settings (RLS) has become apparent with studies showing the limitations of immunological monitoring. We compared the Abbott m2000 Real-Time (Abbott) HIV-1 assay with the Roche AMPLICOR Monitor v1.5 (Roche) HIV-1 assay over a range of VL concentrations. Three hundred and eleven plasma samples were tested, including 164 samples from patients on ART >/= six months and 147 from ART-naive patients. The Roche assay detected >/=400 copies/mL in 158 (50.8%) samples. Of these, Abbott produced 145 (91.8%) detectable results >/=400 copies/mL; 13 (8.2%) samples produced discrepant results. Concordance between the assays for detecting HIV-1 RNA >/=400 copies/mL was 95.8% (298/311). The sensitivity, specificity, positive predictive value and negative predictive value of Abbott to detect HIV-1 RNA >/=400 copies/mL were 91.8%, 100%, 100% and 92.2%, respectively. For the 151 samples with HIV-1 RNA >/=400 copies/mL for both assays, a good linear correlation was found (r = 0.81, P < 0.0001; mean difference, 0.05). The limits of agreement were -0.97 and 1.07 log(10) copies/mL (mean +/- 2 SD). The Abbott assay performed well in our setting, offering an alternative methodology for HIV-1 VL for laboratories with realtime polymerase chain reaction (PCR) capacity.
机译:随着研究显示免疫学监测的局限性,对在资源有限的环境(RLS)中接受抗逆转录病毒疗法(ART)的HIV患者进行病毒载量(VL)监测的需求已变得显而易见。我们在一定的VL浓度范围内,将Abbott m2000实时(Abbott)HIV-1检测方法与Roche AMPLICOR Monitor v1.5(Roche)HIV-1检测方法进行了比较。测试了311个血浆样本,其中包括164个ART> == 6个月患者的血浆样本和147个未接受ART的患者样本。罗氏测定法在158个(50.8%)样品中检测到> / = 400拷贝/ mL。其中,雅培产生的检测结果> / = 400拷贝/ mL 145(91.8%); 13个(8.2%)样品产生了不同的结果。检测HIV-1 RNA> / = 400拷贝/ mL的检测方法之间的一致性为95.8%(298/311)。 Abbott检测HIV-1 RNA> / = 400拷贝/ mL的敏感性,特异性,阳性预测值和阴性预测值分别为91.8%,100%,100%和92.2%。对于这两种测定的151个HIV-1 RNA> / = 400拷贝/ mL的样品,发现良好的线性相关性(r = 0.81,P <0.0001;平均差,0.05)。一致性限制为-0.97和1.07 log(10)个拷贝/ mL(平均+/- 2 SD)。雅培测定在我们的环境中表现良好,为具有实时聚合酶链反应(PCR)能力的实验室提供了HIV-1 VL的替代方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号