A comparison has been made of the reject rates of plain images for three separate periods when film, computed radiography (CR) and PACS systems were in operation throughout the Hammersmith Hospital, London. There was a statistically significant reduction in the overall percentage reject rate across all examinations from 9.9% to 8.1% when the hospital changed from using a conventional film based system to a CR system. There was a further reduction in the reject rate to 7.3% when the hospital moved to a hospital-wide PACS system, but this change was not statistically significant. Using estimations of the total number of images used, the percentage reject rates were 6.6% for film, 5.5% for CR and 5.5% for PACS. Thus, if the radiation dose for each image is unchanged, and the same types of images are used for the examination of each body area, a move from conventional film imaging to phosphor plate imaging provides the potential to reduce the patient population dose.
机译:在伦敦哈默史密斯医院的整个胶片,计算机射线照相(CR)和PACS系统在运行的三个不同时期,对普通图像的拒绝率进行了比较。当医院从使用传统的基于胶片的系统更改为CR系统时,所有检查的总拒绝率从9.9%下降到8.1%,具有统计学上的显着降低。当医院转移到全院PACS系统时,拒绝率进一步降低到7.3%,但是这种变化在统计学上并不显着。使用估计的图像总数,胶卷的废品率分别为6.6%,CR 5.5%和PACS 5.5%。因此,如果每个图像的辐射剂量不变,并且相同类型的图像用于检查每个身体区域,则从传统的胶片成像到磷光板成像的转变提供了减少患者人群剂量的潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号