Patient dose values in a dedicated Greek cardiac centre.

机译:在专用的希腊心脏中心的患者剂量值。

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this study was to collect information on the practice and patient doses in a major Greek cardiac centre, investigate differences between senior cardiologists of various levels of experience and compare results with the literature, in order to optimize angiographic and interventional cardiology procedures. Radiation doses from 292 patients have been studied, 195 of which had undergone coronary angiography and 97 percutaneous transluminal coronary angioplasty. All procedures were undertaken on a Siemens Angioscop X-ray equipment. The system performed under automatic exposure control using pulsed fluoroscopy of 12.5 pulses s(-1) and cine frame rate of 25 frames s(-1). Dose-area product values, fluoroscopy times, total number of cine frames as well as operator's name were collected for each patient. Only senior cardiologists have participated in the study. Median values for dose-area product were 39.1 Gy cm(2) for coronary angiography and 58.3 Gy cm(2) for percutaneous transluminal coronary angioplasty. Median fluoroscopy time was 5.0 min and 9.7 min and median number of frames was 1588 and 1823 for coronary angiography and percutaneous transluminal coronary angioplasty, respectively. Comparison showed that patient dose-area product values were lower than other studies and fluoroscopy time values were comparable. However, the total number of frames used was much higher than other published results. Differences between cardiologists with increased experience have been found. Analysis of the patient dose values obtained initiated a program of radiation protection optimization. The need for continuous training in radiation protection for interventionalists has been verified.
机译:这项研究的目的是在希腊主要的心脏中心收集有关实践和患者剂量的信息,调查具有不同经验水平的高级心脏病专家之间的差异,并将结果与​​文献进行比较,以优化血管造影和介入性心脏病程序。研究了来自292位患者的放射剂量,其中195位接受了冠状动脉造影,而97位经皮腔内冠状动脉成形术。所有程序均在Siemens Angioscop X射线设备上进行。该系统在自动曝光控制下使用12.5脉冲s(-1)的脉冲荧光透视法和25帧s(-1)的电影帧速率执行。收集每位患者的剂量面积乘积值,荧光检查时间,电影帧总数以及操作者姓名。只有高级心脏病专家参加了这项研究。对于冠状动脉造影,剂量-面积乘积的中位数值为39.1 Gy cm(2),对于经皮腔内冠状动脉成形术,中位数值为58.3 Gy cm(2)。对于冠状动脉造影和经皮腔内冠状动脉成形术,荧光检查的中位时间分别为5.0分钟和9.7分钟,中位帧数分别为1588和1823。比较表明,患者剂量-面积乘积值低于其他研究,并且透视时间值可比。但是,使用的帧总数远高于其他已发布的结果。已发现经验丰富的心脏病专家之间存在差异。对获得的患者剂量值的分析启动了辐射防护优化程序。已经证实需要对介入医师进行辐射防护方面的持续培训。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号