A Laboratory Comparison of Real-Time Measurement Methods for 10-100-nm Particle Size Distributions

机译:A Laboratory Comparison of Real-Time Measurement Methods for 10-100-nm Particle Size Distributions

获取原文
获取原文并翻译 | 示例
           

摘要

We present information regarding the relative performance of five TSI particle sizing instruments when presented with several log-normally distributed particle populations that vary in terms of composition, concentration, and modal mean diameter (in the range of 10-100nm) in a controlled laboratory environment. In experiments conducted with NaCl, NaNO3, and organic aerosols, across a total particle concentration suite ranging from approximately 1 x 10(4) cm(-3) to 1 x 10(6) cm(-3), total number concentrations of sub-100-nm diameter particles from four SMPS systems and an FMPS all fall within +/- 50 of each other (and generally are within +/- 30). However, larger discrepancies are evident in the particle size distribution, particularly for the NaCl particles, with an SMPS operated with a water-based CPC exhibiting large negative bias in modal peak concentrations relative to the isobutanol-based SMPS systems and the FMPS. Much closer agreement is found for NaNO3 particles, although the SMPS systems tended to exhibit higher modal peak concentrations, and a slight shifting toward lower modal peak diameter than the FMPS.