首页> 外文期刊>therapeutic drug monitoring >Comparison of Radioimmunoassay with Homogeneous Enzyme Immunoassay for the Determination of Theophylline Concentration in Serum
【24h】

Comparison of Radioimmunoassay with Homogeneous Enzyme Immunoassay for the Determination of Theophylline Concentration in Serum

机译:Comparison of Radioimmunoassay with Homogeneous Enzyme Immunoassay for the Determination of Theophylline Concentration in Serum

获取原文
           

摘要

The homogeneous enzyme immunoassay (EMITreg;) and radioimmunoassay (RIA) procedures for the quantitation of theophyl-line concentrations in serum were evaluated and compared. Both methods exhibited curves linear over the therapeutic range (range of concentration, 2.0ndash;33.0 mu;g/ml). Precision was acceptable for both methods with the coefficient of variation being less than 8.3percnt;. The RIA procedure was found to be slightly more precise. The correlation between the EMIT and RIA procedures was found to be sufficient to allow the procedures to be used interchangeably. This would facilitate the test availability on a 24 hr, 7 day basis.